Aesculap
High frequency surgical device GN060
4.3

Safe operation

Risk of injury and/or malfunction!
Always carry out a function check prior to using
the product.
WARNING
Risk of injury when using the product beyond the
field of view!
Apply the product only under visual control.
WARNING
Switching on the device
In order to be able to switch on the HF device it must be connected to the
mains voltage. The green LED 1 lights up.
Press the "Device on/off" button 3 in order to switch on the HF device.
The green LED 2 lights up.
Directly after switching on the HF device, all symbol LEDs and segments in
the display light up for approx. 1 s. For HF devices with the "automatic
switching" option, the display "AU" appears in the display briefly.
About 1 s after switching on, the HF device has the following basic setting:
"Micro" operating mode
"Foot switch" activation type
Dosage "0.1"
Selecting the operating mode
"Micro-Macro" button 4 - press this button in order to select the oper-
ating mode.
The corresponding LED "Micro" 5 or LED "Macro" 6 lights up.
The operating mode changes every time the button is activated. For
each change of operating mode, the lowest dose is set and shown in
the display.
Select activation type
For HF devices with the "automatic starting" option, the activation type
can be selected.
Press button 19 in order to select the activation type.
The corresponding LED 17 or LED 18 lights up.
The activation type changes every time the button is activated.
Adjust the dosage
The dosage can be changed between the minimum and maximum values
of the selected operating mode:
Press "+" button 10 in order to increase the dosage or "-" button 7 in
order to reduce the dosage.
In the "Micro" operating mode, the value per counting step changes by
0.1, and in the "Macro" operating mode by 1.
If the button is kept pressed, the counting frequency increases after
the third counting step.
The displayed dosage values correspond to the output in watts with nom-
inal load, e.g. if the displayed dosage value "43" corresponds to an output
of 32 watts with a load of 100 ohms, see HF output performance charts.
10
®
The dosage can also be changed by retrieving a stored device adjustment,
see Retrieve stored device settings.
Store device settings
For storing complete device settings including dosage values (empirical
values), there are four memories available. Values that are already present
are overwritten in the event of a storage process.
Press button 20 in order to activate the storage process.
LED 11 lights up.
Press one of the four storage buttons 12 in order to store the current
device settings.
The LED 13 over the pressed storage button lights up when storage has
been carried out successfully.
Store operating example to the device settings
As an example, in the operating mode "Macro", the dosage level "45"
should be stored in the memory "3".
Be sure that the HF device is in ready for operation.
The green LED 1 and LED 2 light up.
Press the buttons represented in the table consecutively:
Button
Message displayed
"01"
"01" to "45"
"45"
"45"
Symbols lighting up
"Macro" 6
"Macro" 6
"Macro" 6 and LED 11
"Macro" 6 and LED 13