Aesculap
High frequency surgical device GN060
7.

Troubleshooting list

Error codes
In the event of an error. one of the error codes detailed in the following table appears in the display 8.
Error code
F1
F2
F3
F4
F5
E1-E6
EF
16
®
Detection
Multiple keys pressed
Membrane key held for longer than
15 s
Foot switch activated for longer
than 30 s
EEPROM writing error
EEPROM reading error
Error in electronics
Microprocessor error
Cause
Several buttons activated at the
same time or object that presses on
several buttons
A key has been activated for more
than 15 s or an object is pressing on
the keyboard
It was not possible to write
EEPROM
It was not possible to read EEPROM see F4
Remedy
Only press one button or, if neces-
sary, remove object from the key-
board. If an error is still displayed,
there is probably a defect present
in the keyboard. Contact Aesculap
Technical Service.
Activate the keys for a shorter
period of time or, if necessary,
remove the object from the key-
board. If an error is still displayed,
there is probably a defect present
in the keyboard. Contact Aesculap
Technical Service.
Release foot switch and activate it
again. If the error message contin-
ues to be displayed even if the foot
switch is not activated, it is possi-
ble that this or the lead cable of the
foot switch defective. Contact
Aesculap Technical Service.
Switch off the HF device and
switch it on again If the error mes-
sage continues to be displayed,
there is an error in the EEPROM.
Contact Aesculap Technical Ser-
vice.
Switch off the HF device and
switch it on again If the error mes-
sage continues to be displayed,
there is a device error. Contact
Aesculap Technical Service.