Maximum peak output voltage (up)
Risk of injury to patients or users caused by inade-
quate accessories (insulation rating)!
Make certain that the accessory voltage rating
WARNING
specified in the product documentation is
higher than maximum peak output voltage for
the intended setting.
Note
The following graph (see Fig. 3) allows the user to judge whether the HF
device or its output setting is suitable for a given accessory (insulation rat-
ing).
350
300
250
200
150
Micro
100
50
0
0
10
Fig. 3
Maximum peak output voltage (up)
3.

Preparation and setup

Non-compliance with the following instructions will preclude all respon-
sibility and liability in this respect on the part of Aesculap.
When setting up and operating the product, adhere to
– national regulations for installation and operation,
– national regulations on fire and explosion protection.
Note
For the safety of patients and users it is essential that the mains power cord
and, especially, the protective earth connection are intact. In many cases
defective or missing protective earth connections are not registered imme-
diately.
Connect the device via the potential equalization terminal at the rear
panel of the device to the potential equalization system of the room
used for medical purposes.
Note
The potential equalization lead can be ordered from the manufacturer as
art. no. GK535 (4 m length) or TA008205 (0.8 m length).
Macro
20
30
40
Output Set
3.1

First use

Risk of injury and/or product malfunction due to
incorrect operation of the electromedical system!
WARNING
3.2

Stacking of units

The stacking provisions integrated in the device cover allows secure
mounting of an accessory device (such as the JET irrigation unit GN090)
on top of the HF device GN060.
Make certain the system is set up on a sufficiently stable support (e.g.
a table, ceiling support, equipment cart, etc.).
Stacking products fitted with stacking devices:
A
50
Fig. 4
Stacking of units
Legend
A Accessory device, e.g. JET irrigation unit GN090
B Locking screw
C Stacking cone
Do not exceed a stack height of 475 mm.
Remove the caps from stacking cones C.
Unscrew the feet of the product that will be on top.
Move stacking cones C to the correct mounting position by applying a
screwdriver on locking screw B and turning it counterclockwise to the
positive stop.
Place the upper device on the one below it.
Securing the device: Turn locking screw B clockwise to the positive
stop.
Slightly lift the units to check that they are safely connected to each
other.
Place the units in a stable position.
Adhere to the instructions for use of any medi-
cal device.
C
B
7