Switching Off
the Pump
Warnings/Alarms
1. Make sure the therapy is stopped and the pump is
in Standby (refer to "Stopping Therapy" on
page 18).
2. If connected, disconnect the mains power.
3. Press and hold the Power button for approximately
2 seconds until the LCD screen goes blank.
NOTE
The Mains indicator above the Power button will
remain green if the pump is connected to the mains
power and extinguished if the mains power is
disconnected.
On detection of a fault condition, the pump provides a
visual-only warning followed by an audible and visual
alarm if the fault is not cleared.
1. The visual-only warning is a fault icon on the LCD
screen. The LED indicators will remain green, but
may begin flashing, and there is no audible alarm.
2. If the fault is not cleared then the warning is
replaced by an audible and visual alarm, which
consists of:
• A fault icon on the LCD screen.
• The LED indicators on the front and underside
of the pump change to yellow.
• An audible alarm will sound.
The warning and alarm can be cleared by either:
• Rectifying the fault on the system, or
• Pressing the Middle Control button to put the pump
into Standby.
Refer to Section 7, Page 25 "Troubleshooting" for the
alarms, their possible causes and their remedies.
19