Condition
Low Pressure/Leak
Kinked/Blocked Tube
Garment Disconnected from the
Pump
Description and Corrective Action
The warning is activated after 4 minutes and shows
a leak in garment 1 or its tubing.
The warning changes to an alarm after an additional
6 minutes (10 minutes total).
Examine the garment and tubing for leaks. The
warning or alarm will be cleared if the leak is
repaired. If the alarm continues, replace the affected
garment.
The warning is activated after 4 minutes and shows
garment 1 has a kinked or blocked tube.
The warning changes to an alarm after an additional
6 minutes (10 minutes total).
Examine the garment and tubing for kinks or
blockages. The warning or alarm will be cleared if
the kink or blockage is repaired. If the alarm
continues, replace the affected garment.
This warning is activated after 1 minute if the pump
detects that a garment has been disconnected while
the pump is in the Run state; it shows that garment 1
has been disconnected. The "garment type" icon
flashes, additionally the green LED indicators will
flash. The therapy timer will be paused, and the
duration the garment has been disconnected from
the pump is displayed in minutes in the bottom right
corner.
The warning changes to an alarm after an additional
9 minutes (10 minutes total) if the pump continues to
detect that a garment has been disconnected while
the pump is in the Run state; it shows that garment 1
has been disconnected.
Reconnect or replace the garment. The warning or
alarm will be cleared if the garment is reconnected.