Electromagnetic Compatibility

Product has been tested for compliance with current regulatory standards
regarding its capacity to block EMI (electromagnetic interference) from external
sources.
Some procedures can help reduce electromagnetic interferences:
• Use only Arjo cables and spare parts to avoid increased emissions or decreased
immunity which can compromise the correct functioning of the equipment.
• Ensure that other devices in patient-monitoring and/or life-support areas comply
to accepted emissions standards.
Wireless communications equipment such as wireless computer
network devices, mobile phones, cordless telephones and their base
stations, walkie-talkies, etc. can affect this equipment and should be kept
at least 1.5m away from the equipment.
Intended Environment: Home Healthcare Environment and Professional
Healthcare facility environment.
Exceptions: HF Surgical Equipment and the RF Shielded room of an ME SYSTEM
for magnetic resonance imaging
Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is
necessary, this equipment and the other equipment should be observed to
verify that they are operating normally.
Guidance and manufacturer's declaration - electromagnetic emissions
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
WARNING
WARNING
Compliance
Group 1
Class A
Class A
Complies
34
Electromagnetic environment - guidance
This equipment uses RF energy only for its
internal functions. Therefore its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
This equipment is suitable for use in all
establishments, other than domestic
establishments and those directly connected to
the public low voltage power supply network that
supplies buildings used for domestic purposes.