5�4 Reusable Collection System:
Serial connection up to 5 jars (chapter F)
See pictures chapter F, page 8.
Note
When using serial connection, the flow rate of the suction source is reduced� Make sure the flow
rate is still adequate for your desired application�
1� Insert required number of jars into jar holder� Remove mechanical overflow protection from
all jars except the front jar�
2� Attach lids and put a reusable serial connector in the vacuum ports of all jars except the front
jar� The tubing shall point towards the previous jar�
3� Connect the tubing of the reusable serial connectors with the patient port of the previous jar�
4� a� Take suction tubing from the vacuum source and pull from below the jar holder through the
hole closest to the front jar�
b� Plug in suction tubing into vacuum port on lid of front jar�
c� Connect desired patient tubing to the patient port on the right lower (= last) jar�
5�5 Reusable Collection System: Change-over valve (chapter G)
See pictures chapter G, page 9.
Note
When using change-over valve, suction is switched from one jar to the other manually�
No reduction in flow rate is to be expected�
1� Insert a jar, having the overflow protection attached, to each side of the change-over valve�
2� a� Connect the left tubing from the change-over valve to the left jar� Do the same for the right
side�
b� Connect suction tubing from the vacuum source to the change-over valve, leading the
suction tubing from below through a hole in the jar holder to the jar�
3� Attach desired patient tubing to the patient ports of the left and right jars�
14 |