6. INSTALLATION

6.1 Setup

CAUTION: Always perform a visual inspection of the device prior to use. Look for signs of wear
and tear, damage, and/or corrosion. Do not attempt to use a damaged light source.
Place device on a stable surface such as a table, counter, stand, etc.
NOTE: Avoid placing device in areas that may be splashed with liquid.
WARNING: DO NOT use in any environment with explosive or flammable gases.
CAUTION: The device has vents located along its bottom and rear structure. DO NOT block airflow
in front of, below, or behind the device.
NOTE: The device requires minimum clearance of 5-inch (13 cm) above and behind for adequate
air ventilation.
CAUTION: The device should be isolated from other devices so heat generated by it shall not
affect other devices and vice versa.
Ensure rear power inlet switch is in the OFF position.
Ensure lamp is properly positioned for use. Open the lamp compartment door, and ensure the
lamp is fully seated into the lamp plug.
Close lamp compartment door ensuring the latch engages.
Connect AC power cord to power inlet on rear panel. Ensure plug is fully seated.
CAUTION: Use only power cords provided with the device.
WARNING: To reduce risk of electric shock, connect power cords of peripheral equipment through
a medical isolation transformer.
NOTE: When using a medical isolation transformer, ensure the transformer is adequately rated to
handle the load.
Connect power cord to AC mains supply.
CAUTION: DO NOT operate this device near strong RF energy equipment.

6.2 Connecting a Light Guide

Rotate device turret until the port matches the type of fiber optic cable being connected. Connect
fiber optic cable to distal end (endoscope or other instrument) then insert cable light guide into
light source turret. Leave slack on cable when installing onto head band (Figure 1).
NOTE: Some resistance is normal when inserting light guide into device turret.
LIT-263 Sunoptic Technologies
Rev. A
Figure 1. Headlight cable positioning. (1) Leave slack
®
Page 10 of 134