→ The Axon-Bus prosthetic system and any damaged Axon-Bus components may only be
opened and/or repaired by certified Ottobock Myo-Service technicians.
CAUTION
Dirtying of the electrical contacts
Risk of injuries due to faulty control or malfunction of the Axon-Bus prosthetic system caused
by an insufficient supply of power to the Axon-Bus components.
→ Ensure that the contacts of the charging receptacle are always clean and free of grease.
→ Clean the electrical contacts of the battery charger and charging receptacle regularly using
cotton swabs and a mild soap solution.
→ Avoid damaging the contact surface coating with pointed or sharp objects.
NOTICE
Improper care of the housing
Damage to the housing through the use of acetone, benzine or similar solvents.
→ Only use a damp cloth and mild soap to clean the housing.
4 Delivery/Storage

4.1 Scope of Delivery

1 piece
757B500 AxonEnergy Integral
1 piece
Lamination Dummy, battery
1 piece
Lamination dummy, charging receptacle
1 piece
Drilling template for charging receptacle
1 piece
Component holder set
1 piece
647G592 Instructions for Use
4.2 Storage
Please see the specifications in the "Technical Data" section.

5 Product description

5.1 Function
The 757B500 AxonEnergy Integral serves to provide power to the Axon-Bus prosthethic system.
6 Components
6.1 Rechargeable Battery
The rechargeable battery (Fig. 1, item 1) consists of Li-ion cells with integrated electronics. This
protects the battery against short circuits, overvoltage, undervoltage and charging outside the
allowable temperature range.
6.2 Charging Receptacle
The charging receptacle (Fig. 1, item 2) has the following functions:
Contacts for charging the battery
LED display to indicate the current battery charge level
LED display for providing feedback on operating states
Button to turn the Axon-Bus prosthetic system on and off, display the battery charge level and
open the Axon-Bus gripping component in an emergency