Battery empty
Battery 50% charged
Battery fully charged
8.4 Safety Shutoff
The purpose of the safety shutoff is to protect the battery; it is triggered in case of:
•
Excessively high or low temperature
•
Overvoltage or undervoltage
•
Short circuit.
After a short circuit the battery charger must be connected and then disconnected in order for the
electronics to be activated.
8.5 Open the Axon-Bus Gripping Component in an Emergency
8.6 Bluetooth Function
Pressing the button in the charging receptacle and holding for more than four seconds with the
Axon-Bus prosthetic system turned off activates the Bluetooth function of the connected Axon
Master 13E500. The blue LED lights up. Once a connection to the PC has been successfully
established, the LED flashes blue.
8.7 Beep Function
Action
Switch on the Axon-Bus prosthet
ic system
Press and hold the button for one to
four seconds.
Switch off the Axon-Bus prosthet
ic system
Press the button and hold for more
than one second.
Charge the Axon-Bus prosthetic
system
Connect charging plug, charging
process starts.
Stop charging
Disconnect charging plug. Charging
ends and the Axon-Bus prosthetic
system is switched off.
Orange LED is illuminated
Yellow LED is illuminated
Green LED is illuminated
This safety function allows the Axon-Bus gripping component to be
opened regardless of the control signals present.
1) With the Axon-Bus prosthetic system switched on, press the char
ging receptacle button and hold for approximately three seconds
until the Axon-Bus gripping component begins to open. A pulsating
beep sounds as it opens.
2) Releasing the button immediately stops the process of opening the
Axon-Bus gripping component and turns the entire Axon-Bus pros
thetic system off.
Beep
1x long
Beep
Beep
2x short
3x short
Beep
Beep
1x short,
pulsating
1x long
Ottobock | 15