Correct torque settings ...
... can be lifesaving in some applications.
For this reason, please note the following
points:
Unless otherwise indicated in the user's
internal regulations (e.g. test equipment
inspection to ISO 9000 et seq), an
inspection must take place after approx.
5000 operations or every 12 months,
whichever is the shorter. The time frame
(12 months) starts with the first usage.
If an inspection shows that there is
excessive deviation, the torque wrench
will have to be readjusted (see page 43).
Additional important safety
points ...
... may apply depending on the
application. These are found in the
appropriate sections marked with danger
symbols ã.
Operation
The Manoskop
instrument and must be treated with the
utmost care. Avoid subjecting the tool to
physical knocks, chemicals or excessive
temperatures beyond the limits given in
these instructions.
ã
C AUTI O N !
In order to ensure the
cut-out is working
accurately, it should be
checked at regular
intervals.
®
730N is a measuring
29