EMC GUIDANCE
The ME EQUIPMENT or ME SYSTEM is suitable for home
healthcare environments.
WARNING: Don't near active HF surgical equipment
and the RF shielded room of an ME system for
magnetic resonance imaging, where the intensity of EM
disturbances is high.
WARNING: Use of this equipment adjacent to or stacked
with other equipment should be avoided because it could
result in improper operation. If such use is necessary, this
equipment and the other equipment should be observed
to verify that they are operating normally.
WARNING: Use of accessories, transducers and
cables other than those specified or provided by the
manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic
immunity of this equipment and result in improper
operation.
WARNING: Portable RF communications equipment
(including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches)
to any part of the equipment MDS7001B, including cables
specified by the manufacturer. Otherwise, degradation of
the performance of this equipment could result.
40
Technical description:
1. All necessary instructions for maintaining BASIC
SAFETY and ESSENTIAL PERFORMANCE
with regard to electromagnetic disturbances
for the excepted service life.
2. Guidance and manufacturer's declaration
electromagnetic emissions and Immunity.
Table 1
Guidance and manufacturer's declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/flicker
Complies
emissions IEC 61000-3-3
Electromagnetic environment –
guidance
The device uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
The device is suitable for use in all
establishments, other than domestic
and those directly connected to the
public low-voltage power supply
network that supplies buildings used
for domestic purposes.