The materials of the cuff have been tested and found to
comply with requirements of ISO 10993-5:2009 and ISO
10993-10:2010.
It will not cause any potential sensation or irritation reaction.
Adapter is specified as a part of ME EQUIPMENT.
If user experiences discomfort during a measurement,
such as pain in the arm or other complaints, press any
button to release the air immediately from the cuff.
Loosen the cuff and remove it from your arm.
If the cuff pressure reaches 40 kPa (300 mmHg), the unit
will automatically deflate. Should the cuff not deflate
when pressure reaches 40 kPa (300 mmHg), detach the
cuff from the arm and press any button to stop inflation.
Before use, make sure the device functions safely and is
in proper working condition. Do not use the device if it is
damaged in any way. The continuous use of a damaged
unit may cause injury, improper results,
or serious danger.
Cleaning
Do not wash the cuff in a washing machine or dishwasher!
Dust in the environment may affect the performance
of the unit. Please use a soft cloth to clean the whole
unit before and after use. Don't use any abrasive or
volatile cleaners.
8
CAUTION:
Please dispose of accessories, detachable parts, and the
ME Equipment according to the local guidelines.
The operator shall not touch the adapter and the
patient simultaneously.
If you have any problems with this device, such as setting
up, maintaining or using, please contact Medline customer
service. Don't attempt to or repair the device by yourself
in the event of malfunctions.
Please report to Medline if any unexpected operation
or events occur.
Keep the unit out of reach of infants, young children
or pets to avoid inhalation or swallowing of small parts.
It can dangerous or even fatal.
Be careful of strangulation hazards due to cables
and hoses, particularly due to excessive length.
ME equipment should warm for at least 30 min. from the
minimum storage temperature between uses until it is
ready for intended use. ME equipment should cool for at
least 30 min. from the maximum storage temperature
between uses until it is ready for intended use.
Wireless communications equipment such as wireless
home network devices, mobile phones, cordless
telephones and their base stations, and walkie-talkies
can affect this equipment and should be kept at least
a distance away from the equipment. The distance is