DBS Lead Implantation
6. Verify that the DBS Lead has not moved from the desired location.
7. Place the split suture sleeve onto the DBS Lead. Be careful not to move the DBS
Lead when placing the suture sleeve.
8. Rotate the suture sleeve until the slit is facing the skull.
Note: If the slit is not located properly, the DBS Lead may not be protected by the suture
sleeve and may be in direct contact with the mini plate.
9. Place a mini plate (not provided) over the suture sleeve and DBS Lead.
CAUTION: Do not place the mini plate in direct contact with the DBS Lead. Ensure
that the suture sleeve lies between the mini plate and the DBS Lead.
10. Secure the mini plate to the skull per the manufacturer's instructions.
Vercise™ DBS Physician Manual
91098825-11
21 of 341