Vercise™ DBS Physician Manual
Selection #2 – Impedances
When the "Impedances" option is selected from the Enter Clinician Options menu, the
Remote Control will take the measurements via telemetry; this will take a few seconds.
Note: During this measurement the serial number of the Stimulator will be displayed on
the screen.
Eventually, the Remote Control will display the Contact Impedance Status screen.
Contacts 1 through 8 (Lead position 1‑L) are represented by the rectangles in the top row;
contacts 9 through 16 (Lead position 2-R) are represented by the rectangles in the bottom
row.
Contacts within the acceptable impedance range are displayed as solid rectangles; high
impedance contacts (above 4500 Ohms) are represented by hollow rectangles.
Pressing any key, except the
of contacts 1 – 8. Subsequent pressing of any key, except the
show values for contacts 9 – 16.
Any key press other than the
which allows for repeating the measurement or exiting this function.
Note: The Remote Control will terminate the CP Ready state and transition to the sleep
mode if there is no IR signal after 15 minutes.
All buttons are active during CP Ready and pressing any button returns the
Remote Control to the Level screen.
Stimulation may be turned on or off during CP Ready.
Once IR Communication is established, the Remote Control will terminate
communication and transition to the sleep mode if there is no IR activity after 15
minutes.
Vercise™ DBS Physician Manual
91098825-11
54 of 341
Stim On/Off button, will display the value of the reading
Stim On/Off button will show the Measure Again screen
Stim On/Off button, will