Using the IPM-400 to Source and Read (cont'd)
Calibrate Pressure Transmitters (cont'd)
7. Press the READ/POWER TRANSMITTER button to toggle between reading milliamps and
supply voltage to power the transmitter.
8. Disconnect both the pressure input and output wires from the two-wire transmitter to be
calibrated.
9. Carefully make all pressure connections.
10. Connect the red lead of the meter to the plus (+) output of the transmitter and the black
lead to the minus (–).
11. Connect a Greenlee IPP-50P hand pump or regulated pressure source to the input of the
two-wire transmitter, as shown. Calibrate the transmitter in the usual manner.
All manuals and user guides at all-guides.com
Pressure
Transmitter
HI
MAX
SET
READ
LO
MIN
SOURCE
READ
STORE
RESET
% DP FLOW
SOURCE READ
mA
V
T/C
RTD
PRESSURE
FREQ
°C / °F
TYPE
ZERO
ENG UNITS
mA / %
READ/POWER
TRANSMITTER
DISPLAY
POWER
IPM-400
37