Safety Precautions Before And During Use - Mectron Piezosurgery touch Manual De Uso

Ocultar thumbs Ver también para Piezosurgery touch:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 142
05.3 Û SAFETY PRECAUTIONS BEFORE AND DURING USE
WARNING: Before starting to operate
with the device, make sure that you
have stock material (handpiece, inserts,
wrenches) available to use in case
EN
the device stops due to a fault or of
inconveniences.
CAUTION: Contraindication. Do not
perform treatments on prosthetic artifacts
made of metal or ceramics. The ultrasonic
vibrations could lead to the de-cementing
of the artifacts.
WARNING: Contraindications. Do not
use the PIEZOSURGERY® touch on patients
who carry heart stimulators (Pace-makers)
or other implantable electronic devices.
This precaution also applies to the
operator.
WARNING: Checking device status
before the treatment. Always check that
there is no water underneath the device.
Before every treatment, always check
that the device works perfectly and that
the accessories are efficient. In case you
uncover operating abnormalities, do
not perform the treatment. Contact an
Authorized Mectron Service Center if the
abnormalities concern the device.
CAUTION: CLEAN function. After
the device is used with aggressive and
non-aggressive solutions, it is necessary
to perform a cleaning cycle on the tubes
and the handpiece with the CLEAN
function (see Cleaning and Sterilization
Manual). If the tubes are not cleaned, the
crystallization of the salts may seriously
damage the device.
CAUTION: Contraindications. Allow
reusable, autoclavable items (the
handpiece, the inserts, the torque wrench,
and any other accessory that can be
sterilized ) to gradually return to room
temperature after steam sterilization and
prior to usage. The cooling process must
not be accelerated.
18
WARNING: Infections control.
First use: All new and repaired accessories
are supplied in NON STERILE conditions.
Before use, and after each treatment, they
must be cleaned and sterilised in strict
compliance with the instructions given in
the Cleaning and Sterilization Manual.
Subsequent uses: After every treatment,
clean and sterilize all the reusable parts
and accessories, following the instructions
provided in the Cleaning and Sterilization
Manual.
CAUTION: The electrical contacts inside
the cord connector must be dry. Before
connecting the handpiece to the device,
make sure that the electrical contacts of
the connector are perfectly dry, especially
after the sterilization cycle in autoclave.
If need be, dry the contacts by blowing
compressed air onto them.
CAUTION: CLEAN function. The
CLEAN function must be used after every
treatment, before starting the cleaning and
sterilization procedures.
WARNING: To ensure that the
handpiece cools down, always activate it
with the irrigation circuit correctly installed
and filled. To fill the irrigation circuit,
always use the PUMP function
WARNING: Treatments that require
irrigation. Always check operation of the
irrigation before and during use. Make sure
the fluid outflows from the insert.
Do not use the device if the irrigation does
not work or if the pump is defective.
WARNING: Only use original Mectron
inserts, accessories, and spare parts.
WARNING: Use of non-original
Mectron inserts: this use entails finite
damage to the handpiece threading,
thus compromising correct operation and
risking to cause harm to the patient.

Publicidad

Tabla de contenido
loading

Tabla de contenido