08 Û TECHNICAL DATA
Device compliant to Dir. 93/42/CEE:
Classification as per EN 60601-1:
EN
Device for intermittent operation:
Power supply voltage:
Max. power absorbed:
Fuses:
Operating frequency:
Power types:
Peristaltic pump capacity:
LED system of the handpiece:
Protections of the APC circuit:
Operating conditions:
Transport and storage conditions:
Peristaltic pump tube:
Weights and sizes:
24
Class II a
I
Applied part type B (handpiece, insert)
IP 20 (device)
IP X8 (foot pedal)
60sec. ON - 30sec. OFF with irrigation
30sec. ON - 120sec. OFF without irrigation(ENDO,PERIO)
100-240 Vac 50/60 Hz
120 VA
Type 5 x 20 mm
T 2AL, 250V
Automatic scan
From 24 KHz to 36 KHz
ENDO
PERIO
SPECIAL
CANCELLOUS
CORTICAL
IMPLANT
Adjustable on the touch screen:
ENDO / PERIO - 7 flow levels: from 0 to 6
(from 0 to approximately 75ml/min)
SPECIAL / CANCELLOUS / CORTICAL / IMPLANT
6 capacity levels: from 1 to 6
(from 8 to approximately 75ml/min)
Light function on AUTO: LED system of the handpiece
switches on as soon as the device starts to operate
and switches off 3 seconds after the foot pedal is
released
Light function on ON: The LED of the handpiece is
always on; after 100 seconds of foot pedal disuse it
switches off by itself and the light function moves into
position AUTO
Light Function on OFF: The LED of the handpiece is
always off.
No handpiece detected
Cord interruption
Insert not tightened correctly or broken
from +10°C to +40°C
Relative humidity from 30% to 75%
Pressure of air P: 800hPa/1060hPa
from -10°C to +70°C
Relative humidity from 10% to 90%
Pressure of air P: 500hPa/1060hPa
It is advisable not to exceed 8 sterilization cycles
3,2 Kg
L -W - H
300 x 250 x 95 mm

Publicidad