3) Caution: Before screwing again the header cap to the sensor body, the
silicone ring must be removed, uncovering the small hole shown in the
following picture:
Please, take care that this step is done correctly, otherwise the header
membrane could be damaged, and will be out of warranty.
4) Screw the header cap to the body of the probe. The excess of electrolyte
will leak out across the small hole and the upper part of the cap. Use a cloth
or blotting paper to clean it.
Important: Ensure to fully screw the header cap.
5) Put the silicone ring to its original position, in this way, the small hole will be
covered again with the silicone ring.