Electromagnetic emissions – Ultravision Battery Recharging Station
The Ultravision battery recharging station is intended for use in the electromagnetic environment specified below. The cus-
tomer or the user of the Ultravision battery recharging station should assure that it is used in such an environment.
Emissions test Compliance
Electromagnetic environ-
ment – guidance
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Electromagnetic immunity – Ultravision Battery Recharging Station
The Ultravision battery recharging station is intended for use in the electromagnetic environment specified below. The cus-
tomer or the user of the Ultravision battery recharging station should assure that it is used in such an environment.
I
test
MMUNITY
Electrostatic discharge
(ESD) IEC 61000-4-2
Electrical fast transient/
burst IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8
NOTE
is the a.c. mains voltage prior to application of the test level.
U
T
32
Compli-
Electromagnetic environment - guidance
ance
Group 1
The Ultravision battery recharging station uses RF energy only for its inter-
nal function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment. RF emissions
Class A
The Ultravision battery recharging station is suitable for use in all estab-
lishments other than domestic and those directly connected to the public
low-voltage power supply network that supplies buildings used for domestic
Class A
purposes.
Complies
IEC 60601test level
± 8 kV contact
± 15 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
100kHz repetition frequency
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
0 % U
; 250/300 cycle
T
30 A/m
Compliance level
Electromagnetic environment
± 8 kV contact
Floors should be wood, concrete or
ceramic tile. If floors are covered with
± 15 kV air
synthetic material, the relative humidity
should be at least 30 %.
± 2 kV for power
Mains power quality should be that of a
supply lines
typical commercial or hospital environ-
± 1 kV for input/
ment.
output lines
± 1 kV line(s) to line(s)
Mains power quality should be that of a
typical commercial or hospital environ-
± 2 kV line(s) to earth
ment.
0 % U
; 250/300 cycle
Mains power quality should be that of a
T
typical commercial or hospital environ-
ment. If the user of the Ultravision battery
recharging station requires continued oper-
ation during power mains interruptions, it is
recommended that the Ultravision battery
recharging station be powered from an
uninterruptible power supply or a battery.
30 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
guidance