Descargar Imprimir esta página

coltene/whaledent BioSonic US100R Guía De Uso página 10

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 29
10
5.4   E N/IEC   6 0601-­‐1-­‐2:2007   S ub-­‐clause   5 .2.2.2   T able   6 :    
Recommended   s eparation   d istances   b etween   p ortable   a nd   m obile   R F   c ommunications   e quipment   a nd  
the   B ioSonic®   U S100R   u nit.  
The   B ioSonic®   U S100R   u nit   i s   i ntended   f or   u se   i n   a n   e lectromagnetic   e nvironment   i n   w hich   r adiated   R F  
disturbances   a re   c ontrolled.   T he   c ustomer   o r   t he   u ser   o f   t he   P roduct   N ame/Model   u nit   c an   h elp   p revent  
electromagnetic   i nterference   b y   m aintaining   a   m inimum   d istance   b etween   t he   p ortable   a nd   m obile   R F  
communications   e quipment   ( transmitters)   a nd   t he   B ioSonic®   U S100R   u nit   a s   r ecommended   b elow,  
according   t o   t he   m aximum   o utput   p ower   o f   t he   c ommunications   e quipment.  
Rated   m aximum   o utput  
power   o f   t ransmitter   i n  
watts   ( W)  
0.01  
0.1  
1.0  
10  
100  
For   t ransmitters   r ated   a t   a   m aximum   o utput   p ower   n ot   l isted   a bove,   t he   r ecommended   s eparation  
distance   d   i n   m eters   ( m)   c an   b e   e stimated   u sing   t he   e quation   a pplicable   t o   t he   f requency   o f   t he  
transmitter,   w here   P   i s   t he   m aximum   o utput   p ower   r ating   o f   t he   t ransmitter   i n   w atts   ( W)   a ccording   t o  
the   t ransmitter   m anufacturer.  
 
NOTE   1         A t   8 0   M Hz   a nd   8 00   M Hz,   t he   s eparation   d istance   f or   t he   h igher   f requency   r ange   a pplies.  
NOTE   2       T hese   g uidelines   m ay   n ot   a pply   i n   a ll   s ituations.   E lectromagnetic   p ropagation   i s   a ffected   b y  
absorption   a nd   r eflection   f rom   s tructures,   o bjects   a nd   p eople.  
 
Separation   d istance   a ccording   t o   f requency   o f   t ransmitter   i n   m eters   ( m)  
150kHz   t o   8 0MHz  
d   =   1 .2√P
.12  
.38  
1.2  
3.8  
12  
80MHz   t o   8 00MHz  
d   =   1 .2√P  
.12  
.38  
1.2  
3.8  
12  
BioSonic Ultrasonic Scaler
800MHz   t o   2 .5GHz  
d   =   2 .3√P  
.23  
.73  
2.3  
7.3  
23  

Publicidad

loading