Owner's Guide
V. Environmental/Storage Conditions
Environmental Conditions
• Intended for indoor use
• Maximum altitude : 2000m (6600 ft.)
• Temperature range:
• Relative humidity:
50% @ 40°C (104°F)
Storage Conditions
The unit must be stored and transported within the
following conditions:
• Temperature range:
-40°C (-40°F)
• Relative humidity:
(non-condensing)
• Atmospheric pressure:
VI. Warnings
• Use cold water only.
• Do not use the ultrasonic scaler without water.
• Do not touch the metal working tip of an activated Ultrasonic
Scaler with bare hands, as it will cause minor burns.
• Do not use ultrasonics for dental procedures involving the
condensation of amalgam.
• To assure optimum performance use only inserts manufactured
by Coltène/Whaledent Inc.
• Do not restrict air flow. Provide adequate ventilation.
• Do not subject unit to shock or impact.
• Do not immerse unit in water.
• This product is intended for use by trained healthcare
professionals only.
• Use this product in accordance with the instructions in the
Owner's Guide.
• Unspecified or improper use of this product may impair safety
protection.
• Do not place the unit on or next to a radiator or other heat
source. Excessive heat may damage the unit components. Place
the unit in well ventilated areas.
• Flush water supply to this device regularly.
• Close the main shut-off valve to the unit before leaving
each day.
• Do not use this device for any type of dental surgery.
• Ground fault interrupt electrical outlets are recommended for
use with any electromechanical devices that are used with or
near water.
• Equipment not suitable for use in the presence of flammable
mixtures.
• Handpiece cannot be sterilized. Use Alpet D2 or BioSonic
Wipeout to disinfect handpiece.
40°C (104°F)
5°C (41°F)
80% up to 31°C (88°F)
%
70°C (158°F)
100%
%
10%
1060hPa
500hPa
VII. Precautions for Ultrasonic Prophylaxis
Procedures
• Ultrasonic inserts "wear out" with use. Inserts that have 2 mm
of wear lose about 50% of their scaling efficiency.
• If excessive wear is evident, or the insert has been bent, discard
the insert immediately. It may break during use.
• Use retraction to protect the tongue, cheek and lips when the
scaler is in use to avoid contact between the two.
VIII. Order Information
Units:
Catalog No.
Voltage
US100R100
100V
US100R115M
115V
US100R115T
115V
US100R230CE
230V
US100R230UK
230V
US100R240
240V
Inserts:
Frequency 25 kHz
Catalog No.
Description
US1025K
#10 Universal
USG1025K
#10 SuperSoft™
US1025KOM
#10 OptiMist™
US1025KSP
#10 Universal Slim
USG1025KSP
#10 Slim SuperSoft
US1025KOS
#10 Slim OptiMist
USG1025KOS
#10 Slim SuperSoft (Optimist)
US100025K
#1000 Universal Triple Bend
Inserts:
Frequency 30 kHz
US1030K
#10 Universal
USG1030K
#10 SuperSoft
US1030KOM
#10 OptiMist
US1030KSP
#10 Universal Slim
USG1030KSP
#10 Slim SuperSoft
US1030KOS
#10 Slim OptiMist
USG1030KOS
#10 Slim SuperSoft (Optimist)
US100030K
#1000 Universal Triple Bend
US10OR (OR-alpha)
Replacement O-Ring Gaskets
US100RXF
Water Filter Replacement Kit
IX. Troubleshooting
Generally, check all lines and connections to and from the unit, a loose
plug or connection will often create problems. Check the settings on
the unit's knobs. Although service and repair of the BioSonic US100R
ultrasonic scaler should be performed by authorized personnel, the
following are some basic trouble shooting procedures that will help
avoid unnecessary service calls.
A. Unit will not operate
1.
Check the ON/OFF switch making sure it is in the ON position,
and that the detachable power cord is fully seated in the
receptacle on back of unit (ON/OFF indicator light should
be on).
2.
Check that the unit's electrical plug is fully seated in an
appropriate AC receptacle, and that the receptacle is active.
B. No water at insert
1.
Assure that the H
O control is properly adjusted. A fine
2
mist or rapid drip of water should be observed at the tip of
the insert.
2.
Check that the main shut-off valve is open.
5
Frequency
Power
50/60 Hz
100W max
50/60 Hz
100W max
50/60 Hz
100W max
50/60 Hz
100W max
50/60 Hz
100W max
50/60 Hz
100W max
Qty
1
1
1
1
1
1
1
1
Qty
1
1
1
1
1
1
1
1
1
1