10. Technical data
Classification acc. to Directive 93/42/EEC
Art.
Designation
no.
OP930
Xenon light source
Model
Voltage range (current
consumption)
Frequency
Safety rating
Device protection
Time-Current characteris-
tic
Breaking capacity
Construction
Weight
Dimensions L x W x H
Applied part
Xenon lamp
EMC
Norms applied
OP930
100 V to 240 V (5.5 A to 2.3 A)
50 Hz to 60 Hz
I
T 4.0 AH / 250 V
T (time lag)
H (1 500 A)
5 mm x 20 mm
6.5 kg
305 mm x 305 mm x 125 mm
Type CF
180 W xenon
Short-arc reflector lamp (14.5 V)
IEC 60601-1-2
IEC 60601-1
10.1 Ambient conditions
Temperature
Class
I
Relative
humidity
Atmospheric
pressure
Note
The atmospheric pressure of 770 hPa corresponds to a maximum operating
altitude of 2 500 m.
11. Disposal
Note
The user institution is obliged to process the product before its disposal, see
Validated reprocessing procedure.
Adhere to national regulations when disposing of or recy-
cling the product, its components and its packaging!
The recycling pass can be downloaded from the Extranet
as a PDF document under the respective article number.
(The recycling pass includes disassembling instructions for
the product, as well as information for proper disposal of
components harmful to the environment.)
Products carrying this symbol are subject to separate col-
lection of electrical and electronic devices. Within the
European Union, disposal is taken care of by the manufac-
turer as a free-of-charge service.
Detailed information concerning the disposal of the product is avail-
able through your national B. Braun/Aesculap agency, see Technical
Service.