Xenon Light Source OP930
1.1
Safe for use in conformance with IEC regulations
Operating environment
Risk of fatal injury from electric shock!
►
Do not open the product.
DANGER
►
Connect the product only to a grounded power
supply.
►
Do not use the xenon light source at more than 2500 m above sea level.
►
Make certain that the ambient temperature does not exceed 35 °C.
►
Make certain that the device's ventilation ducts are not covered.
►
Make certain that the following minimum distances from other devices
are maintained:
– on the sides 50 mm
– from the rear 60 mm
►
Make certain that only non-conductive optical cables are used. This
applies in particular to products from other manufacturers, in order to
meet CF requirements.
►
To avoid failure of the xenon lamp during an operation: Make certain
that the xenon lamp is replaced after 500 hours of operation (i.e., its
guaranteed service life). Check the operating time counter.
1.2
Safety information regarding XBO reflector lamps
Risk of injury from explosion!
►
Exercise caution when working with the xenon
lamp. The permissible inner pressure of the
WARNING
xenon lamp when cold is ca. 20 bar.
►
Do not use the xenon lamp more than 1.25
times longer than its average service life. During
operation, the inner pressure of the lamp
increases threefold.
Risk of injury to eyes from strong ultraviolet radi-
ation and high luminance!
►
Do not look directly into or at the xenon lamp.
WARNING
►
Make certain that xenon lamps are operated only with enclosed xenon
light sources.
►
Only operate XBO lamps with the safety cladding and protecting cap.
►
When performing maintenance or repairs, always wear protective garb
(safety glasses and gloves).
►
When doing maintenance or repair work, make certain that the reflec-
tor opening is not pointing towards any persons.
4
®
2.
Product description
2.1
Scope of supply
Note
The xenon light source OP930 does not have a spare lamp. As the xenon
lamp may fail before the guaranteed operating period has elapsed, always
have a spare lamp available, see Accessories/Spare parts.
Art. no.
Designation
OP930
Xenon light source
OP935
Aesculap-Storz optical cable connection
TA009789
Instructions for use for Xenon Light Source OP930
TA022130
Notes on electromagnetic compatibility (EMC)
2.2
Components required for operation
Art. no.
Designation
OP930
Xenon light source
TE730 to TE780
Power cord
OP906
Optical cable
OP913
OP914
2.3
Intended use
The xenon light source OP930 is used to illuminate the operative field and
body cavities during medical applications.