Aesculap
Xenon Light Source OP930
4.
Working with the xenon light source
4.1
System set-up
Risk of burns from excessively high operating tem-
peratures!
WARNING
The xenon lamp, optical cable socket and optical
cable connections get hot.
►
Do not situate optical cables near any flamma-
ble objects (e.g., drape).
Danger of blinding from light rays!
►
Never look into the open end of an optical cable
or endoscope that is connected.
WARNING
►
Set up the device on an even surface in a non-sterile area.
►
Make certain that the device support (e.g., table, hanging ceiling light,
medical cart or similar) is sufficiently stable.
►
Make certain that the device is safeguarded against vibration during
operation.
Connecting the accessories
Risk of injury due to unapproved configuration
using additional components!
►
For all applied components, ensure that their
DANGER
classification matches that of the application
component (Type CF) of the respective device.
Combinations of accessories that are not mentioned in the present
instructions for use may only be employed if they are specifically intended
for the respective application, and if they do not compromise the perfor-
mance and safety characteristics of the products.
Also note that any equipment connected at the interfaces must demon-
strably meet the respective IEC standards (e.g. IEC 60950 for data process-
ing equipment, IEC/DIN EN 60601-1 for electromedical devices).
All configurations must comply with basic standard IEC/DIN EN 60601-1.
Any individual connecting devices with one another is responsible for such
configuration and must ensure compliance with basic standard IEC/DIN
EN 60601-1 or applicable national standards.
►
Please address your B. Braun/Aesculap partner or Aesculap Technical
Service with any inquiries in this respect; for a contact address, see
Technical Service.
Connecting the endoscope
►
Connect the endoscope to the instrument connection 12.
6
®
Connecting the optical cable
The endoscope must be connected to the instrument connection 12.
►
Insert the optical cable 11 into the optical cable connection 6 until it
snaps into place.
►
Do not crush or crimp the optical cable 11 or wind it too tightly (min-
imum diameter 150 mm).
Keep the xenon light source separated from the optical cable
►
Separate the xenon light source from the optical cable after use.
►
In order to prevent the xenon light source from slipping, hold the xenon
light source firmly in one hand and pull out the optical cable 11 with
the other hand.
Changing the optical cable connection
The Xenon Light Source OP930 is delivered with a Aesculap-Storz fiber-
optic connection. This can be exchanged for the fiber-optic connection
systems Wolf (article no. OP936), Olympus (article no. OP837) and ACMI
(article no. OP938), see Accessories/Spare parts.
Risk of fatal injury from electric shock!
►
Before dismantling the fiber-optic connection
system, disconnect the power plug.
DANGER
►
Do not insert anything into the device opening.
Risk of burns!
►
Only replace the fiber-optic connection system
DANGER
after the device has completely cooled down.
►
Do not insert anything into the device opening.
►
Do not operate the device without a fiber-optic
connection system being installed.
If a fiber-optic cable is plugged in, remove it:
►
Rotate the optical cable connection 6 in a counterclockwise direction
and then remove it.
►
Mount the other optical cable connection 6 by rotating it as far as it
will go in a clockwise direction.
Always ensure that the fiber-optic connection is firmly connected with
the housing.