Remove battery 23 from the battery compartment
by knocking the unit against the palm of the hand,
see Fig. 8.
Fig. 8
Removing the battery
Intraoperative battery change
Risk of injury from hot bat-
tery!
Battery may be hot after use in
WARNING
the machine.
Note
The function checks must be carried out prior to each
surgical application of the unit and after each intraop-
erative battery change!
Remove the battery with
the battery removal device
and let it cool.
The battery removal device is used for intraoperative
battery changes ensuring that sterile conditions are
safely maintained.
Hold dermatome 1 with the battery compartment
upward and remove cover 21.
Mount sterile battery removal device 25, see Fig. 9.
25
Fig. 9
Using the battery removal device
Slightly shake dermatome 1 with battery removal
device 25 mounted and the battery compartment
facing downward.
Rechargeable battery 23 will slip smoothly into
battery removal aid 25.
Hand battery removal device 25 containing the dis-
charged battery 23 to an unsterile helper.
Insert charged battery 23, see Inserting the battery.