Descargar Imprimir esta página

B.Braun Aesculap Acculan 3Ti Instrucciones De Manejo/Descripción Técnica página 21

Ocultar thumbs Ver también para Aesculap Acculan 3Ti:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 70
Mechanical alkaline cleaning and thermal disinfection
Machine type: single-chamber cleaning/disinfecting machine without ultrasound
Phase
Step
I
Pre-rinse
II
Cleaning
III
Intermediate
rinse
IV
Thermal disin-
fection
V
Drying
DW:
Drinking water
FD–W:
Fully desalinated water (demineralized, microbiological, at least of drinking water quality)
*Recommended: BBraun Helimatic Cleaner alcaline
Insert the product in its proper position in the
ECCOS fixation. Insert the battery compartment
over the clamp and swivel device into the ECCOS
fixation so that the trigger 4 points upwards.
Ensure that the blade cover 6 is kept open through
the entire processing procedure.
Dismantle the flap rod 18 for cleaning. Store flap
rod 18, flaps 12 and nut 2 separately in tray.
After automatic cleaning/disinfection, check visible
surfaces for residues and repeat the cleaning/disin-
fection process as needed.
D
t
Water
[°C/°F]
[min]
quality
<25/77
3
DW
55/131
10
FD–W
> 10/50
1
FD–W
90/194
5
FD–W
-
-
-
Chemistry/Note
-
Concentrate, alkaline:
– pH ~ 13
– <5 % anionic surfactant
working solution 0.5%
– pH ~ 11*
-
-
mind. 10 min at max. 120 °C
6.10 Inspection, maintenance and
checks
Note
Aesculap recommends lubricating the flap rod, flaps,
sliding surfaces and other moving parts, (e.g. trigger,
cover lids) with STERILIT-Power-Systems oil spray
GB600 or the STERILIT-Power Systems drip lubricator
GA059 prior to every sterilization. Remove excess oil
with a lint-free wipe.
Allow the product to cool down to room tempera-
ture.
Inspect the product after each cleaning and disin-
fecting cycle to be sure it is: clean, functional, and
undamaged.
Check the product for any damage, abnormal run-
ning noise, overheating or excessive vibration.
Set aside the product if it is damaged.
en
19

Publicidad

loading

Este manual también es adecuado para:

Ga670