1.
General information
1.1
Intended use
Task/Function
The dermatome GA670, combined with dermatome
blade, is used for split skin removal in adjustable depth.
Application Environment
The product fulfills the requirements for type BF pursu-
ant to IEC/DIN EN 60601-1 and is used in operating
rooms in sterile environments of explosion risk areas
(such as areas with pure oxygen or anesthesia gases).
1.2
Main functions and design charac-
teristics
Oscillation fre-
min. 0 min
quency
Operating
Operation with non-periodic load
mode
and speed changes (type S9 pur-
suant to IEC EN 60034-1)
Electrical systems generally heat up during continual
operation. It is advised to give the system a break after
use to cool down, as listed in the table on operating
mode.
Heating depends on the tool used and the load. After a
certain number of repetitions, the system should cool
down. This procedure prevents the system overheating
as well as possible injury to the patient or user.
The user is responsible for the use and adherence to the
pause sequence described.
1.3
Indications
The type and area of application depend on the tool
selected.
4
-1
to max. 6 500 min
60 second application,
60 second pause
10 repetitions
30 min cooling time
Max. Temperature 48 °C
1.4
Absolute contraindications
The product is not licensed for use on the central ner-
vous system or central circulatory system.
1.5
Relative contraindications
The safe and effective use of the product greatly
depends on influences which can only be controlled by
the user. Therefore the specifications provided repre-
sent framework conditions only.
Clinically successful use of the product is dependent on
the knowledge and experience of the surgeon. The sur-
geon must decide which structures it is sensible to
treat and take into account the safety and warning
information contained in these instructions for use.
2.
Safe handling
-1
WARNING
WARNING
Remove the transport packaging and clean the new
product, either manually or mechanically, prior to
its initial sterilization.
Prior to use, check that the product is in good work-
ing order.
See "Information on electromagnetic compatibility
(EMC) for Acculan components" TA022450; see
Aesculap Extranet at https://extranet.bbraun.com
Risk of injury and material
damage if this product is not
used as intended!
Use the product only for its
intended purpose.
Risk of injury and damage to
property due to improper han-
dling of the product!
Follow the instructions for
use of all products used.