en
Stage VI
►
Rinse disinfected surfaces after the prescribed
reaction time for at least 1 minute under running
demineralized water.
►
Drain any remaining water fully.
6.9
Automatic cleaning/disinfection with manual pre-cleaning
Note
The cleaning and disinfection device must fundamen-
tally have a tested efficacy (such as FDA approval or CE
label pursuant to DIN EN ISO 15883).
Manual pre-cleaning with brush
Phase
Step
I
Rinsing
II
Brushes
DW:
Drinking water
RT:
Room temperature
►
Do not clean the product in a ultrasonic bath and
do not immerse the product in any fluids. Let any
fluid incursions drain out immediately, otherwise
there is a danger of corrosion and loss of function.
Phase I
►
Dismantle the flap rod 18 for cleaning. Store flap
rod 18, flaps 12 and nut 2 separately in tray.
►
Mobilize non-rigid components during cleaning.
►
Thoroughly clean the product under running water.
Phase II
►
Mobilize non-rigid components during cleaning.
►
Brush difficult to access surfaces with a suitable
plastic cleaning brush for at least 1 min.
►
After manual preparation, check visible surfaces for
residue and repeat the pre-cleaning process as
needed.
Note
For details on difficult to access surfaces, see Acculan
pre-cleaning and care information TA016000 (available
in Aesculap Extranet at https://extranet.bbraun.com).
18
D
t
Conc.
[°C/°F]
[min]
[%]
RT (cold)
-
-
RT (cold)
-
-
Stage VII
►
Dry the product in the drying phase with suitable
equipment (such as lint-free cloths, pressurized
air).
Note
The washer-disinfector machine used for processing
must be serviced and checked at regular intervals.
Water
Chemistry/Note
quality
DW
until visually clean
DW
until visually clean