Descargar Imprimir esta página

B.Braun Aesculap Acculan 3Ti Instrucciones De Manejo/Descripción Técnica página 17

Ocultar thumbs Ver también para Aesculap Acculan 3Ti:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 70
6.2
General notes
Dried or affixed surgical residues can make cleaning
more difficult or ineffective and lead to corrosion.
Therefore, no more than 6 hours should be left between
use and preparation, pre-clean temperatures >45 °C
liable to fusing should not be employed and disinfec-
tants liable to fusing (aldehyde or alcohol-based)
should not be used.
Excessive neutralizing agents or basic cleaners may
result in a chemical attack and/or fading and the laser
marking becoming unreadable either visually or by
machine in case of non stainless steel.
On stainless steel, residues containing chlorine or chlo-
ride (such as surgical residues, drugs, saline solutions in
water for cleaning, disinfection and sterilization) may
lead to corrosion (pitting corrosion, tensile corrosion)
and thus to the destruction of the product. These must
be removed by rinsing thoroughly with demineralized
water and then drying.
Perform additional drying, if necessary.
Only process chemicals that have been tested and
approved (e.g. VAH or FDA approval or CE mark) and
which are compatible with the product's materials
according to the chemical manufacturers' recommen-
dations may be used for reprocessing the product. All
the chemical manufacturer's application specifications
must be strictly observed. Failure to do so can result in
the following problems:
Visual material changes (such as fading or color
changes in titanium or aluminum). For aluminum,
the application/process solution only needs to be
pH >8 to cause visible surface changes.
Material damage (such as corrosion, cracks, breaks,
premature aging or swelling).
Do not use metal cleaning brushes or other abra-
sives that would damage the product surface and
could cause corrosion.
Further detailed advice on hygienically safe and
material-/value-preserving reprocessing can be
found at www.a-k-i.org, section Publications, Red
Brochure – Proper maintenance of instruments.
6.3
Preparation at the place of use
Remove all attached components from the product
(dermatome blade, battery, accessories).
Remove any visible surgical residues as much as
possible with a damp, lint-free cloth.
Place the dry product in a sealed waste container
and forward it on for cleaning and disinfection
within 6 hours.
6.4
Dismantling prior to carrying out
the reprocessing procedure
Remove battery 23 from the product 1, see
Removing the battery.
Open the clamping levers 7 to remove the derma-
tome blade 24, see Removing the dermatome
blade.
Disassemble the product 1 immediately after use,
according to instructions.
Removing the flap rod of the dermatome
Loosen nut 2 by turning it clockwise (left-handed
thread).
Screw the nut 2 back to the end of the visible sur-
face.
Press on nut 2 and push flap rod 18 approx. 4 mm
to one side.
Twist flap rod 18 until it can be removed.
Remove flap rod 18.
Slide off the flaps 12 from flap rod 18.
The flap rod 18 has been dismantled.
6.5
Preparation before cleaning
Prior to the first mechanical cleaning/disinfection
process, mount the ECCOS fixations in a suitable
tray (e.g. JC254R) or use the ECCOS tray GB256R
with fixations equipped.
Insert the product in its proper position in the
ECCOS fixations, see Fig. 11.
en
15

Publicidad

loading

Este manual también es adecuado para:

Ga670