9. Technical specifications
Test voltages
Maximum resistance reading
Short circuit current
Test voltages accuracy
Megohmmeter basic
accuracy
Advanced features
Built-in printer
Serial data output
Built-in chronometer
Environmental protection
Safety class
E.M.C
Electrostatic immunity
Electro magnetic irradiation
immunity
Power supply
Battery charger
Operating temperature range
: 100 V - 250 V - 500 V - 1.000 V directly, one
button selectable. 100 V to 1.000 V in 50 V
steps. DC, negative.
: 1 TΩ @ 1 kV
: 1.5 ± 0.5 mA
: ± 3 % of nominal test voltages @ R = 10 GΩ
: 5 % of reading ± 3 digits
(1 MΩ to 500 GΩ at any test voltage)
: Automated Polarization Index computing
Automated Dielectric Absorption Ratio
computing
: Prints elapsed time, actual voltage and
resistance measured each 15 seconds (offered
as an optional feature in MD-1035e).
: RS232 @ 4800 bps. Suitable for data collection
in an external serial printer, computer or data-
logger.
: Shows elapsed time in mm:ss format. Count
starts automatically for each measurement.
: IP54 with closed lid.
: Meets the requirements of IEC 61010-1/1990,
IEC 61010-1/1992 amendment 2
: In accordance with IEC 61326-1
: In accordance with IEC 61000-4-2
: In accordance with IEC 61000-4-3
: Internal rechargeable 12 V - 2.3 Ah battery or
100 - 240 V~ mains
: 100 - 240 V~ mains
: -5 °C to 50 °C