EN-24
RS232 Command Set
ID <xxxx>
ID
MODE
START_SHAKE
PAUSE_SHAKE
STOP_SHAKE
TARGET_SPEED <xxxx>
TARGET_SPEED
MEASURED_SPEED
TIMER
TIMER 0
TIMER <hh>:<mm>:<ss>
TIMER RESET
LOCK
UNLOCK
PARAM <xxxx>
Set ID
1 ≤ xxxx ≤ 9999
Read ID value
Read mode:
0: idle
1: shake
2: paused
99: error
Start shake function and timer
Pause shake function and timer
Stop the shake function; reset timer
Set the target speed value
Read target speed value
Read measured speed value
Read current timer value
Set Timer Mode = Count Up
Set Timer Mode = Count Down;
Timer = hh:mm:ss
Reset timer to set value; timer continues counting.
No change to shake function.
Lock user interface
Unlock user interface
Enable output of comma separated list of parameters, formatted as follows:
<TIMER [hh:mm:ss]>, <ID>, <MODE>, <TARGET SPEED>, <MEASURED
SPEED>, <ERROR CODE>,
Output can be periodic, where xxxx represents the desired time interval:
xxxx = 0
Single output
1 ≤ xxxx ≤ 9999
Periodic output every xxxx seconds