3.2.4.3.2 Probe set up
Before installing the probe, it is necessary to fill the cap with electrolyte. Please,
follow carefully the following steps:
1) Unscrew the header cap from the sensor body and keep both parts in a
clean and stable surface.
Important: Don't touch the membrane and the gold electrode, they could
be contaminated.
2) Fill the header cap with the supplied electrolyte. Fill it gently to avoid
creating bubbles.
3) Caution: Before screwing again the header cap to the sensor body, the
silicone ring must be removed, uncovering the small hole shown in the
following picture:
Please, take care that this step is done correctly, otherwise the header
membrane could be damaged, and will be out of warranty.
Manual BSONE-EVO v.2
63 / 95