Brown wire (Supply): Connect to #14 label.
Green wire (Measure): Connect to #13 label.
White wire (Temperature): Connect to #8 label.
Black wire (GND): Connect to #12 label.

3.2.6.3 Operation and adjustment

Once the conductivity sensor is connected, the chlorinator will detect it
automatically.
The main screen will show the measure, and also any warning related with the lack
or too much salt conditions:
BSONE-EVO units
Lack of salt: Measure < 4g/l
Too much salt: Measure > 8g/l

3.2.6.3 Calibration and maintenance

The conductivity sensor comes already calibrated, however, it is posible to Access
to a calibration menu in case that a measurement fine tuning would be required.
Use an external measure system and then set the value in the salt adjust menú.
Press OK to confirm the new calibration.
This sensor does not require any special maintenance. However, if an incorrect
value is observed, an visual inspection of the measure head is recommended, in
order to detect any strange body which could be interfering with the correct
operation.