Dräger Polytron 8000 Serie Instrucciones De Uso página 41

Ocultar thumbs Ver también para Polytron 8000 Serie:

Publicidad

Idiomas disponibles

Idiomas disponibles

Menu
Default set-
ting
Response
normal
behavior
Display cap-
0
ture range off-
set
Display cap-
-750 ppm
ture range
(Methane)
low
-315 ppm
(Propane)
-405 ppm
(Ethylene)
Display cap-
750 ppm
ture range
(Methane)
high
315 ppm
(Propane)
405 ppm
(Ethylene)
1) The maximum limit of the beam block depends on the type of gas
detector and the measured gas selected.
Type 340
Menu
Default set-
ting
Test gas
Propane
Test gas units %LEL
Test gas con-
50 %LEL
centrations
Alarm A1
20 %LEL
Alarm A"
40 %LEL
LEL type
NIOSH
Measured
Propane
gas
Units of mea-
%LEL
sured gas
Full scale
100 %LEL
value
Beam block
Off
Beam block
2 mA
current
Beam block
2.5 %LEL
limit
Adjustment
360
interval
[days]
Response
normal
behavior
Instructions for use
|
Dräger Polytron
Range
normal / fast
The absolute values of the
display capture range must
be within these limits:
-1000 to 2200 ppm
(Methane type 334)
-850 to 850 ppm
(Propane type 334)
-1200 to 1150 ppm
(Ethylene type 334)
Range
0.01 to 100 Vol% with %LEL
= 0.3 to 100 %LEL
IEC / PTB / NIOSH / configu-
rable
Vol% / PPM / %LEL / %LIE
20 to 100 %LEL
On/Off
0 to 3.5 mA
1)
0 to max. 15 %LEL
0 to 720
normal / fast
®
8000 Series
Menu
Default set-
ting
Display cap-
0
ture range off-
set
Display cap-
-750 ppm
ture range
(Methane)
low
-85 ppm
(Propane)
Display cap-
750 ppm
ture range
(Methane)
high
85 ppm
(Propane)
1) The maximum limit of the beam block depends on the type of gas
detector and the measured gas selected.
14.3.8 Polytron 8720 IR
Menu
Default set-
ting
Test gas
CO
2
Test gas units Vol%
Test gas con-
4.0 Vol%
centrations
Alarm A1
1 Vol%
Alarm A2
2 Vol%
LEL type
-
Measured gas CO
2
Units of mea-
Vol%
sured gas
Full scale
10 Vol%
value
Beam block
Off
Beam block
2 mA
current
Beam block
0.1 Vol%
limit
Adjustment
360
interval
[days]
Response
normal
behavior
Display cap-
340 ppm
ture range off-
set
Display cap-
-200 ppm
ture range low
Display cap-
200 ppm
ture range
high
Factory default settings
|
Range
The absolute values of the
display capture range must
be within these limits:
-850 to 425 ppm
(Propane type 340)
-1800 to 2200 ppm
(Methane type 340)
Range
0.2 to 100 Vol%
0.01 to 100 Vol%
Vol% / ppm
0.2 to 100 Vol%
On/Off
0 to 3.5 mA
0 to 0.1 Vol%
0 to 720
normal / fast
The absolute values of the
display capture range must
be within these limits:
-1000 to 1000 ppm
(carbon dioxide)
en
41