6.2 Bendable argon electrode
Bendable argon electrodes can be oriented to the specific operation site.
6.3 Recommended settings
Select the gas flow according to the selected argon electrode and the
indication.
Mode
Monopolar Cut - Argon
Monopolar Coag - Argon
28
BOWA-IFU-MN030-088-ARGON-HANDLES-AND-ELECTRODES-S6-ML-20230910
WARNING
Risk of injury from distally protruding needle
When changing from coagulation to the cutting electrode,
slide the shaft tube carefully.
NOTE
Risk of damage to the device from excessive bending
Do not exceed the maximum overall bending radius of 90° with
bendable
argon electrodes (REF 932-060, 932-061, 932-062).
WARNING
Risk of injury from excessive intra-abdominal pressure
Take intra-abdominal pressure into account for laparoscopic
use.
Effect
1 - 9
---
Max. 90°
Max. power
Flow rate
80 – 100 W
4.0 - 8.0 l/min
60 – 80 W
2.0 - 4.0 l/min