Precautions
- Contact between the file and the gums (this may cause a false
reading indicating that the apex has been reached).
- Use of an ultrasonic scaler with the counter electrode attached to
the patient (electrical noise from the scaler could interfere with the
apex localization).
- Use of the apex locator in conjunction with an electric scalpel.
- Use of a damaged lip clip, file clip or fork.
For apex localization, concentrations of NaOCl higher than 5 %
may result in reduced accuracy.
As a safety precaution in order to avoid over-instrumentation, it is
recommended to proceed as follows: place the file onto an
endodontic ruler, where the Propex Pixi
'0.0'. Subtract a minimum of 0.5 mm from the measured file length.
Please also respect the following precautions:
- For your own safety, be aware of wearing personal protective
equipment (gloves, glasses, mask).
- If the bar graph makes sudden large movements in the coronal
part of the canal, slowly continue advancing the file toward the
apex until the signal returns to normal.
- This Propex Pixi
used in combination with any other apparatus or system. It must
not be used as an integral component of any other apparatus or
system. Using replacement parts or accessories not supplied by
the original manufacturer or vender could adversely affect the
EMC performance of the Propex Pixi
®
Apex Locator unit must not be connected to or
®
Apex Locator indicates
®
Apex Locator.
11/1346