Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The Product is intended for use in the electromagnetic environment specified below.
The customer or the user of the Product should assure that it is used in such an
environment.
Immunity test
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Electrical fast
transients /
bursts,
IEC 61000-4-4
Surges
IEC 61000-4-5
IEC 60601-1-2
Test level
± 8 kV contact
± 15 kV air
± 2 kV for
power supply
lines
± 1 kV for
input/ output
lines
± 1 kV Line-to-
line
± 2 kV Line-to-
ground
Compliance
level
± 8 kV contact
± 15 kV air
± 2 kV for
power supply
lines
Not Applicable
± 1 kV Line-to-
line
± 2 kV Line-to-