Maintenance, Cleaning, Disinfection and Sterilization
7.2
Disinfection and Sterilization Procedure for Lip Clip,
File Clip and Fork
Foreword
For hygiene and sanitary safety purposes, the lip clip, the file clip and
the fork must be cleaned, disinfected and sterilized before each usage
to prevent any cross-contamination between patients. This concerns
the first use as well as the subsequent uses.
General recommendations
Use only a disinfecting solution which is approved for its efficacy
(VAH/DGHM-listing, CE marking, FDA approval) and in accordance
with the DFU of the disinfecting solution manufacturer. For all metal
instruments, it is recommended to use anticorrosion disinfecting
and cleaning agents.
For your own safety, please wear personal protective equipment
(gloves, glasses, mask).
The user is responsible for the sterility of the product for the first
cycle and each further usage as well as for the usage of damaged
or dirty instruments where applicable after sterility.
Limitations and restrictions on reprocessing:
the appearance of defects such as cracks, deformations (bent,
twisted), corrosion, loss of colour coding or marking, are indications
that the devices are not able to fulfil the intended use with the
required safety level.
Use only clean water in all cleaning and rinsing steps.