2.2.3
Medit Scan for Clinics User Guide
Please refer to the User Guide of Medit Scan for Clinics: Medit Scan for
Clinics > Menu > User Guide.
3 Maintenance
CAUTION
▪
E quipment maintenance should only be carried out by a Medit
employee or Medit-Certified company or personnel.
▪
I n general, users are not required to perform maintenance work
on the i700 system besides calibration, cleaning, and sterilization.
Preventive inspections and other regular maintenance are not
required.
3.1 Calibration
Periodic calibration is required to produce precise 3D models.
You should perform calibration when:
▪
T he quality of the 3D model is not reliable or accurate when
compared to previous results.
▪
Environmental conditions such as temperature have changed.
▪
T he calibration period has expired. You can set the calibration
period in the Menu > Settings > Calibration Period (Days).
12
User Guide
T he calibration panel is a delicate component. Do not touch
the panel directly. Check the calibration panel if the calibration
process is not performed properly. If the calibration panel is
contaminated, please contact your service provider.
W e recommend performing a calibration periodically. You can
set the calibration period via Menu > Settings > Calibration
Period (Days). The default calibration period is 14 days.
3.1.1
How to Calibrate i700
①
Turn on the i700 and launch the Medit Scan for Clinics.
②
Run the Calibration Wizard from Menu > Settings > Calibration.
③
Prepare the calibration tool and the i700 handpiece.
④
Turn the dial of the calibration tool to position
⑤
Put the i700 handpiece into the calibration tool.
⑥
Click "Next" to start the calibration process.
⑦
W hen the calibration tool is mounted properly in the correct
position
, the system will automatically acquire the data.
1
⑧
W hen data acquisition is complete at position
dial to the next position.
⑨
R epeat the steps for positions
position.
⑩
W hen data acquisition is complete at the
will automatically calculate and show the calibration results.
.
1
, turn the
1
–
and the
2
8
LAST
position, the system
LAST