5 Electro-Magnetic Compatibility Information
5.1 Electromagnetic Emissions
The i700 system is intended for use in the electromagnetic
environment as specified below. The customer or the user of the i700
system should ensure that it is used in such an environment.
Guidance and Manufacturer's Declaration – Electromagnetic Emission
Emission Test
Compliance
RF Emissions
Group 1
CISPR 11
RF Emissions
Class A
CISPR 11
Harmonic Emissions
Class A
IEC 61000-3-2
Voltage Fluctuations
Complies
/ Flicker Emissions
WARNING
This i700 system is intended for use by healthcare professionals only.
This equipment/system may cause radio interference or may disrupt
the operation of nearby equipment. It may be necessary to take
mitigation measures, such as re-orienting or relocating the i700 or
shielding the location.
20
User Guide
5.2 Electromagnetic Immunity
Electromagnetic
Environment – Guidance
The i700 uses RF energy only for
its internal function. Therefore,
its RF emissions are very low
and are not likely to cause any
interference in nearby electronic
equipment.
The i700 is suitable for use in all
establishments. This includes
domestic establishments and
those directly connected to the
public low-voltage power supply
network that supplies buildings
used for domestic purposes.
Guidance 1
T he i700 system is intended for use in the electromagnetic
environment as specified below. The customer or the user of
the i700 system should ensure that it is used in such an
environment.
Guidance and Manufacturer's Declaration – Electromagnetic Emission
Immunity
IEC 60601
Compliance
Test
Test Level
Level
Electrostatic
± 8 kV contact
± 8 kV contact
Discharge
± 2 kV, ± 4 kV,
± 2 kV, ± 4 kV,
(ESD) IEC
± 8 kV, ± 15 kV
± 8 kV, ± 15 kV
61000-4-2
air
air
Electrical Fast
±2 kV for power
±2 kV for power
Transient
supply lines
supply lines
/ Burst IEC
±1 kV for input/
±1 kV for input/
61000-4-4
output lines
output lines
Electromagnetic
Environment –
Guidance
Floors should be
made of wood,
concrete, or ceramic
tiles. If floors are
covered with a
synthetic material,
relative humidity
of at least 30% is
recommended.
The mains power
quality should be
that of a typical
commercial
or hospital
environment.