Radiated RF
3 V/m
3 V/m
IEC 61000-4-3
80 MHz
to 2.7 GHz
▪ NOTE 1 : At 80 MHz and 800 MHz, the higher frequency range applies.
▪ N OTE 2 : T hese guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from
structures, objects, and people.
▪ N OTE 3 : T he ISM (Industrial, Scientific, and Medical) bands between 150 kHz
and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz;
26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
Guidance 4
T he i700 system is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled.
Portable RF communications equipment should be used no
closer than 30 cm (12 inches) to any part of the i700 system.
Otherwise, degradation of the performance of this equipment
could result.
d is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey should be less than
the compliance level in each
frequency range.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Communications
IEC 61000-4-3
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
Immunity
Band
Service
Modulation
1)
1)
Test
380 –
390
TETRA 400
Modulation
MHz
430 –
FM ±5 kHz
GMRS 460;
470
Deviation
FRS 460
MHz
1 kHz sine
704 –
LTE Band
787
Modulation
13, 17
MHz
217 Hz
GSM 800:900;
Proximity
800 –
TETRA 800;
Fields from
960
iDEN 820;
Modulation
RF Wireless
MHz
CDMA 850;
LTE Band 5
GSM 1800;CDMA
1700 –
1900;GSM 1900;
1990
Modulation
DECT;LTE Band
MHz
217 Hz
1, 3, 4, 25;UMTS
Bluetooth;WLAN
2400 –
802.11b/g/n;
2570
Modulation
RFID 2450;
MHz
217 Hz
LTE Band 7
5100 –
WLAN
5800
Modulation
802.11a/n
MHz
217 Hz
IEC 60601
Compliance
Test Level
Level
Pulse
27 V/m
27 V/m
18 Hz
28 V/m
28 V/m
Pulse
9 V/m
9 V/m
Pulse
28 V/m
28 V/m
18 Hz
Pulse
28 V/m
28 V/m
Pulse
28 V/m
28 V/m
Pulse
9 V/m
9 V/m
Intraoral Scanner i700