Medit i700 Guia Del Usuario página 23

Escáner 3d intraoral destinado a registrar digitalmente las características topográficas de los dientes y los tejidos circundantes.
Ocultar thumbs Ver también para i700:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 195
±0.5 kV, ±1 kV
±0.5 kV, ±1 kV
differential mode
differential mode
Surge IEC
61000-4-5
±0.5 kV, ±1 kV,
±0.5 kV, ±1 kV,
±2 kV common
±2 kV common
mode
mode
0% Uт (100% dip
0% Uт (100% dip
Voltage
in Uт) for
in Uт) for
Dips, Short
0.5/1 cycles
0.5/1 cycles
Interruptions,
70% Uт (30% dip
70% Uт (30% dip
and Voltage
in Uт) for
in Uт) for
Variations on
25/30 cycles
25/30 cycles
Power Supply
0% Uт (100% dip
0% Uт (100% dip
Input Lines
in Uт) for
in Uт) for
IEC 61000-4-11
250/300 cycles
250/300 cycles
The mains power
Power
quality should be
Frequency
that of a typical
Magnetic
commercial
Fields
or hospital
(50/60Hz)
environment.
IEC 61000-4-8
The mains power
quality should be
that of a typical
commercial
or hospital
Proximity
environment. If the
Magnetic
user of the i700
Fields in the
system requires
Frequency
continued operation
Range 9 kHz
during power mains
to 13.56 MHz
interruptions, it is
Immunity
recommended that
IEC 61000-4-39
the i700 system
be powered from
an uninterruptible
power supply or a
battery.
NOTE : Uт is the main voltage (AC) prior to the application of the test level.
Power frequency
magnetic fields
should be at levels
characteristic of
30 A/m
30 A/m
a location in a
typical commercial
or hospital
environment.
8 A/m
8 A/m
30 kHz CW
30 kHz CW
modulation
modulation
Resistance to
magnetic fields was
65 A/m
65 A/m
tested and applied
134.2 kHz
134.2 kHz
only to surfaces
PM 2.1 kHz
PM 2.1 kHz
of enclosures
or accessories
7.5 A/m
7.5 A/m
accessible during
13.56 MHz
13.56 MHz
intended use.
PM 50 kHz
PM 50 kHz
Intraoral Scanner i700
21