U se a sterilized and non-abrasive cloth to dry the mirror and the
tip gently.
CAUTION
▪
T he mirror found in the tip is a delicate optical component that
should be handled with care to ensure optimal scan quality. Be
careful not to scratch or smudge it as any damage or blemishes
may affect the data acquired.
▪
M ake sure to always wrap the tip before autoclaving. If you
autoclave an exposed tip, it will cause stains on the mirror,
which cannot be removed. Check the autoclave manual for
more information.
▪
T ips that have been cleaned, disinfected, and sterilized must
remain sterile until they are used on the patient.
▪
M edit is not responsible for any damage, such as distortion of
the tip, that occurs during cleaning, disinfection, or sterilization
operations that are not following the guidelines above.
3.2.3
Mirror
The presence of impurities or smudges on the tip mirror may lead to
poor scan quality and an overall poor scanning experience. In such
situations, clean the mirror following the steps below:
①
Disconnect the scanner tip from the i700 handpiece.
②
P our alcohol on a clean cloth or cotton-tipped swab and wipe
the mirror. Make sure to use alcohol that is free of impurities or
it may stain the mirror. You can use either ethanol or propanol
(ethyl-/propyl alcohol).
14
User Guide
③
Wipe the mirror dry using a dry, lint-free cloth.
④
M ake sure the mirror is free of dust and fibers. Repeat the
cleaning process as necessary.
3.2.4
Handpiece
After treatment, clean and disinfect all other surfaces of the i700
handpiece except for the scanner front (optical window) and end (air
vent hole). Cleaning and disinfecting must be done with the device
turned off. Use the device only after it is completely dry.
The recommended cleaning and disinfecting solution is denatured
alcohol (ethyl alcohol or ethanol) – typically 60 – 70% Alc/Vol.
The general cleaning and disinfecting procedures are as follows :
①
Turn off the device using the power button.
②
Unplug all the cables from the power hub.
③
Clean the filter on the front end of the i700 handpiece.
If alcohol is poured directly into the filter, it may seep
»
inside the i700 handpiece and cause a malfunction.
Do not clean the filter by pouring alcohol or cleaning
»
solution directly into the filter. The filter must be gently
wiped with a cotton or soft cloth moistened with alcohol.
Do not wipe by hand or apply excessive force.
Medit is not responsible for any damage or malfunction
»
that occurs during cleaning that does not follow the
guidelines above.
④
P ut the cover on the front of the i700 handpiece after cleaning
the filter.