Guidance 2
Recommended Separation Distances Between Portable and Mobile
Communication Equipment and the i700
Separation Distance According to the Frequency of
Rated
Maximum
Output
IEC 60601‐1‐2:2014
Power of
the
150 kHz to 80 MHz
Transmitter
d = 1.2 √P
[W]
0.01
0.12
0.1
0.38
1
1.2
10
3.8
100
12
For transmitters rated at a maximum output power not listed above, the recommended
separation distance (d) in meters (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
▪NOTE 1 : A t 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
▪NOTE 2 : T hese guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects, and people.
▪
Guidance 3
T he i700 system is intended for use in the electromagnetic
environment specified below. The customer or the user of the
22
User Guide
Transmitter [M]
80 MHz to 2.7 GHz
d = 2.0 √P
0.20
0.63
2.0
6.3
20
i700 system should ensure that it is used in such an
environment.
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
IEC
Immunity
60601
Compliance
Test
Test
level
Level
3 Vrms
150 kHz to
80 MHz
3 Vrms
Outside
ISM Bands
amateur
Conducted
RF IEC
61000-4-6
6 Vrms
150 kHz to
80 MHz
6 Vrms
In ISM Bands
amateur
Electromagnetic
Environment – Guidance
Portable and mobile RF
communications equipment
should not be used closer to
any part of the Ultrasound
System, including cables, than
the recommended separation
distance. This is calculated using
the equation applicable to the
frequency of the transmitter.
Recommended Separation
Distance (d):
d = 1.2 √P
IEC 60601-1-2:2007
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 80 MHz to 2.5 GHz
IEC 60601-1-2:2014
d = 2.0 √P 80 MHz to 2.7 GHz
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer,