Replacement of osmotic membranes: 
To replace osmotic membranes follow instructions below: 
 
 Close the water infeed and outfeed valves of the main body. 
 Close the cock on the storage tank. 
 Remove the upper protection of the main body. 
 Discharge the internal pressure by opening a utility. 
 
 
 Disconnect  hoses  from  quick  couplers  on  the  caps  of  the  vessels 
containing the membranes. Contain any water spills with paper or cloths. 
 Unscrew the caps of vessels and remove the exhausted membranes . 
 
 Fit the new membranes into the vessels, by fully pushing them 
 Close the vessels by deeply screwing the caps. 
 Reconnect hoses to quick couplers. 
 Re‐open the water infeed and outfeed valves of the main body. 
 Open the cock on the storage tank. 
 Switch  on  the  main  body,  pick  water  from  utilities  and  be  sure  of  the 
absence of water leak through couplers and caps of the vessels  . 
 Switch off the machine and reposition the upper protection . 
 Flush the membranes for at least five minutes. 
 
N.B.: Exhausted membranes must be disposed of as special waste.  
A. 
Peristaltic pump hose
B. 
Fixing bushings
C. 
Peristaltic pump hose support 
D. 
Non‐return valves
make reference to Fig. 14  above: 
 