Starting:
To start the machine follow instructions below:
Position the tank discharge valve lever (see Fig. 12) to "Open" (lever in
horizontal position).
Switch On Biox through the switch (see Fig. 13). The green LED on the
front panel illuminates.
Switch on the machine through the switch (see Fig. 12). Make sure the
discharge solenoid valve light is ON; otherwise switch it ON.
The machine starts and produces purified water. The pressure gauge on
the front panel shows a pressure of about 6 bars; the "Flow detection"
yellow LED flashes on the Biox simultaneously to the injection of chlorine
dioxide into water.
When the storage tank is full the Machine stops automatically.
Water consumption through the supplied utilities of the practice causes
alternating automatic restarts and stops aimed at keeping the correct
amount of water into the tank.
Dosing amount check:
To check for selection of the dosing amount of chlorine dioxide into water
follow instructions below :
The Machine On but not running, switch the Biox off and then on.
For about two seconds after the restart the "remaining time Display"
indicates the dosing amount" (see Fig. 13) by illuminating one or two LEDs,
then it shows back the remaining solution validity time.
See Tab. 4.8.1 in the "Installation" section of this Manual to see the
current parameter relating to the dosing amount.
37
ENGLISH