12
Manufacturer's guide and declaration – electromagnetic immunity
This camera is intended for use in the electromagnetic environment specified below. The user must ensure that it is used in this envi-
ronment.
Immunity test
IEC 60601
Level of gravity
Electrostatic discharges
± 6 kV in Contact
EN 61000-4-2
± 8 kV in air
Rapid transient peaks
± 2 kV for power
EN 61000-4-4
lines
± 1 kV for input/
output lines
Electric shocks
Differential mode
EN 61000-4-5
± 1 kV
Shared mode
± 2 kV
Power dips, power
• <5% UT –
failures, short power inter-
for 10 ms
ruptions and variations in
• 40% UT –
power voltage
for 100 ms
EN 61000-4-11
• 70% UT –
for 500 ms
• <5% UT –
for 5 s
System frequency
3 A/m
magnetic field (50/60
Hz) IEC61000-4-8
ELECTROMAGNETIC
COMPATIBILITY
Compliance
level
± 6 kV
± 8 kV
± 2 kV
± 1 kV
± 1 kV
± 2 kV
<5% UT
10 ms
<40% UT
100 ms
<70% UT
500 ms
<5% UT
5 s
3 A/m
Electromagnetic environment - guide
The floor must be wooden, concrete or tiled. If the floor is cove-
red with a synthetic material, the relative humidity must be at
least 30%.
The quality of the main power supply must be that of a commer-
cial or typical hospital environment.
The quality of the electric power supply must be that of a com-
mercial or typical hospital environment.
The quality of the main power supply must be that of a commer-
cial or typical hospital environment.
If the user of this camera requires it to continue functioning du-
ring interruptions of the main power supply, it is advisable for
this camera to be powered by an inverter or a battery.
The system frequency magnetic field must be at a level that is
characteristic of location in a commercial or typical hospital en-
vironment.
- 31 -