RANGES AND MEASURES (According to ATS/ERS 2005)
Measurement range (BTPS)
Flow
Volume
Flow resistance 14 l/s (cmH
Measurement accuracy (BPTS)
Volume
Flow
PEF
Temporal accuracy
Volume resolution
Frequency of sampling
Transducer life
511-D00-MUM · Rev 1.01
Fleisch
0 / l/s)
<1,4
2
10% or 300 ml/s (ATS/ERS 2005)
10% or 170 ml/s (EN ISO 23747:2015)
<10ml
100Hz
1400 disinfect.
or 3 years
DATOSPIR
Turbine
Lilly Disposable
0 to ±16 l/s
0 to 10 l
<0,6
(whichever is greater)
3% or 50 ml
5% or 200 ml/s
0,50%
<10ml
100Hz
1400 disinfect.
or 3 years
(Expir. 3 years)
· User's Manual
<1,2
<10ml
100Hz
Only one use
EN 23