is intended for use in the electromagnetic environment specified below. The costumer or the user of
Below mentioned tests are applicable to enclosure. Additionally, test EN-IEC 61000-4-2 is applicable to input/output signal and patient connections.
Immunity test
EN-IEC 60601 test level
Electrostatic discharge
±8 kV contact
(ESD)
±2, ±4, ±8, ±15 kV,
EN-IEC 61000-4-2
Radiated RF
3 V/m
EN-IEC 61000-4-3
80 MHz to 2.7 GHz
80% AM at 1kHz
Proximity fields from RF
(View following table).
communications
EN-IEC 61000-4-3
Power frequency (50 / 60
30 A/m
Hz) magnetic field
EN-IEC 61000-4-8
Frecuency
Band
(MHz)
(MHz)
385
380-390
450
430-470
710
704-787
745
780
810
800-960
870
930
1720
1700-1990
1845
1970
2450
2400-2570
5240
5100-5800
5500
5785
511-D00-MUM · Rev 1.01
Guidance and manufacturer's declaration – electromagnetic immunity
Compliance level
Electromagnetic environment – guidance
±8 kV contact
air
±2, ±4, ±8, ±15
kV,
air
3 V/m
Portable and mobile RF communications equipment should be used no closer than 30cm to any part of
DATOSPIR
where P is the output power rating of the transmitter in watts (W), d is the recommended separation in
meters (m) and E is the immunity level in V/m.
30 A/m
Only applicable to devices sensitive to magnetic fields.
Proximity fields EN-IEC 61000-4-3
Modulation
Power
(W)
Pulso 18Hz
1.8
FM
2
D:+/-5kHz
1kHz sinus
Pulso 217Hz
0.2
Pulso 18Hz
2
Pulso 217Hz
2
Pulso 217Hz
2
Pulso 217Hz
0.2
DATOSPIR
DATOSPIR
should assure that it is used in such an environment.
, including cables. The distance is calculated according to:
Distance
E
(m)
(V/m)
0.3
27
0.3
28
0.3
9
0.3
28
0.3
28
0.3
28
0.3
9
· User's Manual
EN 27